产品中心

Product Center

9180100

产品简介

Flow cell assembly for use with 1 NPT pH Sensors on CLF10sc and CLT10sc Chlorine Analyzers

Exclusive Self Diagnostics
No Reagent Replacement, No Waste Stream
Real-Time Process Control
Compatible with

产品型号:
更新时间:2025-03-10
厂商性质:生产厂家
访问量:732
详细介绍在线留言

Exclusive Self Diagnostics

The CLF10 sc and CLT10 sc analyzers leverage Hach's exclusive self diagnostics to alert users when the process has changed or the instrument needs servicing. Diagnostic features include the Cal Watch algorithm for warning of pH and chlorine calibration deviation and a non-contacting flow sensor for notification of insufficient sample flow.

No Reagent Replacement, No Waste Stream

Chlorine measurement with an amperometric analyzer, such as the CLF10 sc or CLT10 sc, does not require reagents, eliminating the need for routine reagent replacement and waste stream management.

Real-Time Process Control

The CLF10 sc and CLT10 sc analyzers allow for real-time control of disinfection processes by providing continuous readings that indicate when treatment conditions have changed.

Compatible with Hach's Plug and Play Digital Controllers

The CLF10 sc and CLT10 sc analyzers can be used with any Hach sc digital controller. Whether you're measuring turbidity or chlorine, you only need to learn one controller for all your water analysis measurement points. Hach sc controllers have no complicated wiring or setup procedures. Just plug in any Hach digital sensor and it's ready to use without software configuration.

EPA Compliant According to Method 334.0

In accordance with EPA Method 334.0, the CLF10 sc and CLT10 sc analyzers can be used for reporting chlorine residual measurements. Additionally, Hach has created a suite of laboratory products and methods to help with startup and quality control procedures required in Method 334.0. (See page 4 for a partial listing of accessories.)

Specifications

Accuracy:Free Chlorine: ± 3 % of the reference test** (DPD) at constant pH less than 7.2 (±0.2 pH unit)

Accuracy 2:Free Chlorine: ± 10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)

Accuracy 3:Total Chlorine: ± 10 % of the reference test* (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)

Accuracy 4:Total Chlorine: ± 20 % of the reference test** (DPD) at stable pH greater than 8.5

Cable Connection:5 pin, M12 connector

Cable Length:1 m (between gateway and sc controller)

Calibration Method:1-point or 2-point (zero and slope) calibration

Certifications:EMC
CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits)
- EMC Immunity EN 61326-1 (Industrial limits)
Additional certifications when connected to an sc controller

Connection Drain Line:1/2 in OD

Connections:1/4 in OD

Controller Compatibility:sc controller platform

Diameter:Sensor: 1 in ( 25 mm )

Dimensions (H x W x D):5.9 in x 19.5 in x 19 in (151 mm x 495 mm x 483 mm)

Interferences:Free chlorine: monochloramine, chlorine dioxide, ozone, and chalk deposits. Total chlorine: chlorine dioxide, ozone and chalk deposits

Length:Sensor: 7.7 in ( 195 mm )

Lower Limit of Detection (LOD):30 ppb (0.03 ppm)

Material:Corrosion-resistant materials, fully-submersible 
(stainless steel, PVC, silicon rubber and polycarbonate)

Measurement Method:Reagentless, electrochemical, three-electrode amperometric system

Measurement range:0 to 20 ppm

Mounting:Panel or Wall Mounting

Operating Temperature Range:0 to 45 °C

Power Requirements (Voltage):12 V DC

Pressure Limit:0.5 bar, no pressure impulses and/or vibrations

Repeatability:30 ppb or 3%, whichever is greater

Resolution:0.001 ppm (1 ppb)

Response Time:Free Chlorine: 140 seconds or less for 90% change (T90) at a stable temperature and pH

Sample Flow Rate:30 to 50 L/h (7.9 to 13.2 gal/hour), Optimal is 40 L/hour (10.5 gal/hour)

Sample Temperature:5 to 45 °C (41 to 113 °F), compensated for fluctuations

Sampling Time:Continuous

Storage Conditions:0 to 50 °C Store chlorine sensor dry (without electrolyte)

Temperature Compensation:Temperature compensation provided by internal temperature sensor

Warranty:1 Year

Weight:12.13 lbs. (5.5 kg)