产品分类The Hach Surface Scatter 7 sc Turbidimeter (SS7) is uniquely designed so that the light source and photocell never come in contact with the sample. In fluids with high loads of suspended solids this makes sample cell cleaning and replacement unnecessary.
All wetted parts of the Hach SS7 turbidimeter are made with corrosion-resistant materials for extended life. The photo-detector and light source assemblies are protected from the effects of corrosive vapors and heated samples. The SS7 turbidimeter is warranted against defects in materials or workmanship for two years from the date of shipment.
Surface Scatter 7 sc Turbidimeter—Use the SS7 to establish and maintain optimum process control and for reliable monitoring.
• Drinking water influent
• Wastewater effluent
• Flocculation and sedimentation
• Industrial process water
• Food processing waste containing starch, fat, or oil
Surface Scatter 7 sc High Sample Temperature (HST) Turbidimeter—This is an SS7 turbidimeter designed to monitor samples with temperatures of up to 70°C. An innovative moist air removal system is useful where a difference between the sample temperature and the ambient temperature causes condensation and fogging. For use in applications such as:
• Fluids ranging from raw water influent to corrosive white liquor
• High temperature samples—up to 70°C
• Corrosive white, black, and green liquor in paper mills
• Produced water in oil fields containing oil and hydrogen sulfide
• Wherever temperature differences cause condensation and fogging
The SS7 turbidimeter can reliably measure turbidity from 0 to 9999 NTU in samples that vary from clear water to corrosive and high temperature paper mill and oil field samples.
Consistent with its long-standing reputation for quality and customer service, Hach Company warrants all SS7 turbidimeters against defects in materials or workmanship for two years from the date of shipment.
Hach sc200 Controller, Mounting Hardware, sc200 Basic User Manual, SS7sc, Wall Mount Kit, Cylinder Brush, Light Source Alignment Template, ¾ x ¾ NPT Nipple, Calibration Cup, ¾ NPT Ball Valve, SS7sc User Manual, and 500 mL 4000 NTU Formazin Standard. (The controller power cord is ordered separay.)
Accuracy:±5% of reading or ±0.1 NTU (whichever is greater) from 0 to 2000 NTU
Accuracy 2:± 10% of reading from 2000 to 9999 NTU
Alarm:Two nephelometric set-point alarms (SOM required)
Ambient Temperature:0 to 50 °C
Dimensions (H x W x D):25.3 in x 26.6 in x 7.5 in (642 mm x 675 mm x 190 mm)
Enclosure Construction:Plastic instrument enclosure
Enclosure Rating:NEMA 12
Humidity:5 to 95 % non-condensing
Installation category:Category II
Measuring Range:0 to 9999 Nephelometric Turbidity Units (NTU)
Mounting:Wall mount
Power Requirements (Hz):50/60 Hz
Power Requirements (Voltage):100/230 V AC
Repeatability:1.0 % or ±0.04 NTU, whichever is greater
Resolution:0.1 NTU between 100 to 9999.9 NTU between 100 to 9999.9 NTU
Response Time:Initial response in 45 seconds
Sample Flow Rate:Sample flow required 1.0 to 2.0 L/min (15 to 30 gal/hr)
Sample Temperature:(32 to 122°F) 0 to 50 °C (32 to 122°F)
Weight:SS7: 35.05 lbs. (15.9 kg)
Weight 2:SS7/HST: 18 kg (40 lbs.)